• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY? XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941816300
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon deflation failure occurred.The more than 70% stenosed target lesion was located in a moderately to severely tortuous and non-calcified coronary artery.A 3.00 x 16mm synergy xd drug-eluting stent was advanced for treatment.Balloon inflation was performed using a non-boston scientific inflation device, with a 50/50 saline-contrast mixture.After stent deployment, the stent balloon failed to deflate, despite multiple attempts to pull negative pressure.The physician had to remove the stent delivery system, guidewire, guide extender, and guide catheter as a single unit.The procedure was completed with no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY? XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22816646
Report Number2124215-2025-55905
Device Sequence Number8756452
Product Code NIQ
UDI-Device Identifier08714729980964
UDI-Public(01)08714729980964(17)260605(10)34233522
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/05/2026
Device Model NumberH7493941816300
Device Catalogue NumberH7493941816300
Device Lot Number0034233522
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2025
Initial Report FDA Received Date08/15/2025
Date Device Manufactured06/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-