• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SMHX400
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent primary breast augmentation with a 400cc mentor memorygel xtra breast implant on the right breast.Post-operatively, the patient was diagnosed, via physical examination, with right breast capsular contracture (baker grade iii).At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
 
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On august 18, 2025, mentor received additional information indicating that the patient has been scheduled for breast implant removal surgery on (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: capsular contracture.
 
Additional Manufacturer Narrative
On september 29, 2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.On october 8, 2025, mentor received additional information indicating that the patient was also diagnosed with right breast mispositioning.The patient's implants were replaced with two mentor gel implants.
 
Additional Manufacturer Narrative
On (b)(6) 2025, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22817028
Report Number1645337-2025-09046
Device Sequence Number18395782
Product Code FTR
UDI-Device Identifier10081317027857
UDI-Public(01)10081317027857(17)291009(10)9998158
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSMHX400
Device Lot Number9998158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/23/2025
Supplement Date Received by Manufacturer08/18/2025
09/29/2025
10/29/2025
Initial Report FDA Received Date08/15/2025
Supplement Report FDA Received Date09/10/2025
10/15/2025
11/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-