MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0725
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the single-use sphincterotome's cutting wire broke during an ercp (endoscopic retrograde cholangiopancreatography procedure).There were no reports of patient harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.A likely factor causing tears of the coated portion of the cutting wire might be the following.However, the exact cause could not be determined.·it is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22817244
Report Number9614641-2025-01380
Device Sequence Number12928219
Product Code KNS
UDI-Device Identifier04953170184031
UDI-Public04953170184031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0725
Device Lot Number48K30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer08/27/2025
Initial Report FDA Received Date08/15/2025
Supplement Report FDA Received Date09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-