Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the trial liner 52+4 was found to be missing the screw retaining clip used without clip.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the threaded insert was found disassembled and the snap ring was not returned for evaluation.Additionally, the device exhibits an overall worn appearance consistent with constant usage.No other defects were observed.The wear detected on the device may have caused the ring and threaded insert to become loose and ultimately detached.This condition was identified as an end-of-life indicator.The observed condition appears to be consistent with the effects of repeated use and servicing over a period exceeding 20 years.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pin trl +4 neut 520dx36id would have contributed to the complained device issue.Based on the investigation findings, the root cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code pg0605 provided is not a valid finished goods lot number.H11 additional narrative: added: d6b, d10, h4.Corrected: h3.
|