(b)(4).Date sent: 08/15/2025.D4: batch #713a31.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ats45nk device was received with no apparent damage and with a tr45w reload loaded in the device.The reload was received partially fired 1/2 and with the reload lockout spring damaged.The returned device was tested for functionality with test reload and it fired and formed all the staples as intended.The staple line was complete, the staples met the staple form release criteria and the device closed as expected.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If the re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.The event of would not close could not be confirmed as the device closed without any difficulties noted during the functional testing.Although it could not be determine the cause of the reported incident, proper usage of the endo linear cutters - ets45mm is important to ensure functionality.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished ats45nk, with lot/batch number x94p0f, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 713a31, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|