| Catalog Number |
367921 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6) hospital.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® sodium fluoride/potassium oxalate 5 mg/4 mg, the customer noticed that the tube underfilled.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary : bd had not received any samples or photos for investigation.Therefor a total of 100 retained samples were visually inspected, with no issues found.A functional test was performed on 10 retained samples, and all tubes drew within specification limits with no issues identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® sodium fluoride/potassium oxalate 5mg/4mg, the customer noticed that the tube underfilled.There was no health impact or consequence reported.
|
| |
|