MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH BERLIN FACILITY ELECTROSURGICAL GENERATOR "ESG-410"; WA91307C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH BERLIN FACILITY ELECTROSURGICAL GENERATOR "ESG-410"; WA91307C Back to Search Results
Model Number WA91307C
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported during a therapeutic transurethral resection of a bladder tumor (turbt) procedure, the electrosurgical generator had no energy.Both the wired footswitch and wireless footswitch were attempted.The footswitch indicator on the device was indicating the footswitches were passing a signal but there was no energy present.There were no error codes present.The patient was under general anesthesia at that time.The case was subsequently cancelled due to the event.The customer received a replacement electrosurgical generator.The patient was then brought back for a second attempt procedure on (b)(6) 2025, in which they underwent general anesthesia again.The second procedure was completed successfully.There were no reports of patient harm.It was reported that no testing of the device was done before the first procedure attempt.The customer plugged things in as usual and began the case.All appeared normal until the physician was in the patient and the electrosurgical generator would not fire.It did not respond to pedal steps or resets.It did not respond to new loops.All connections were checked and reset without success.Thus, the case was cancelled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTROSURGICAL GENERATOR "ESG-410"
Common Device Name
WA91307C
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
MDR Report Key22819402
Report Number2429304-2025-00226
Device Sequence Number14887655
Product Code GEI
UDI-Device Identifier04042761086349
UDI-Public04042761086349
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/15/2025
Report Date (Section F) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberWA91307C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/18/2025
Event Location Hospital
Date Report to Manufacturer07/18/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/15/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WA22751S HF-RESECTION ELECTRODE "PLASMAROLLER"; WIRED FOOTSWITCH; WIRELESS FOOTSWITCH
Patient SexUnknown
Date Report Sent to FDA08/15/2025
-
-