EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
| Model Number |
9600TFX20A |
| Medical Device Problem Codes |
Degraded (1153); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Per the instruction for use (ifu), valve stenosis is a potential risk associated with bioprosthetic heart valves.Per the valve academic research consortium (varc), valve stenosis can result from a number of factors, including pannus, calcification, support structure deformation (out-of-round configuration), trauma (cardia-pulmonary resuscitation, blunt chest trauma), endocarditis, prosthetic valve thrombosis, and native leaflet prolapse impeding prosthetic leaflet motion.Valve stenosis may result in symptoms such as shortness of breath and decreased exercise tolerance, which may be accompanied by an increased gradient across the valve.Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention, or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.Due to the limited information provided, the cause of the stenosis is unknown.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|
| |
|
Event or Problem Description
|
|
As reported by the field clinical specialist, approximately 8 years post transfemoral tavr procedure with a 20 mm sapien 3 (s3) valve, the valve is failing due to stenosis.The patient received a 20mm sapein 3 ultra resilia valve inside the 20 mm s3 valve.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
As reported by the field clinical specialist (fcs), approximately 8 years post transfemoral tavr procedure with a 20 mm sapien 3 (s3) valve, the valve is failing due to stenosis.The patient received a 20mm sapien 3 ultra resilia valve inside the 20 mm s3 valve.No patient complications were reported.Based on feedback from the fcs, the perceived root cause of the stenosis is degeneration.The gradient was 63/35 mmhg, and the valve area/index was 0.9 cm2.Thickened cusps were observed.The patient was discharged following the valve-in-valve procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for additional information received.Sections b.4, g.3, g.6, and h.2 have been updated.Corrections have been made to b.7, h.6: clinical code, device code, investigation findings and h.11.An addition was made to b.5.Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with the use of bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.Investigation results suggest/indicate patient factors (advanced age, hypertension, renal dysfunction (nephrolithiasis), and prior radiation therapy) may have caused or contributed to the degeneration.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|
| |
|