| Model Number |
VTICM5_12.1 (-9.50/2.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
| Date of Event |
12/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Iop elevation from baseline, secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced elevated iop.In a separate visit the lens was explanted and on the same day different surgery a replacement lens of a longer length was implanted.The issue was resolved with the replacement lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information.H6- method code 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work orders(s), including the suspected device, have been manufactured within established parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue claim# (b)(4).
|
| |
|