| Catalog Number |
BNST3210 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: weight unknown / not provided.A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #45 was removed due to infection.Symptoms as a result of the event: pain.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 3331 and 4110.H10: narrative/data was updated.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigation, no malfunction occurred upon investigation.The reported event remains non-verifiable as it is a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for the infection of an implant have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
|
| |
|
Search Alerts/Recalls
|