• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NCOMPASS NITINOL TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NCOMPASS NITINOL TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G32701
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
G4 ¿ pma/510(k) #: exempt.H3: device evaluation is anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported the ncompass nitinol tipless stone extractor¿s basket would not open prior to use in retrograde intrarenal surgery (rirs).The issue was found when the device was opened.The device did not make patient contact.The procedure was completed by using another same-type device.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 annex g, d4: udi.E3: occupation unknown.Investigation ¿ evaluation.It was reported the ncompass nitinol tipless stone extractor¿s basket would not open prior to use in retrograde intrarenal surgery (rirs).The issue was found when the device was opened.The device did not make patient contact.The procedure was completed by using another same-type device.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.Functional tests and visual inspection of the returned complaint device was also conducted.One device was returned for investigation with no original packaging.Handle does not actuate basket formation.Handle disassembled, does not manually actuate.Photos of returned device show damage to the basket sheath near the yellow support sheath.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for this incident.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, cook has concluded that the device was manufactured to specification and that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: precautions.Do not use excessive force to manipulate this device.Damage to the device may occur.Based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCOMPASS NITINOL TIPLESS STONE EXTRACTOR
Common Device Name
FFL, DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key22824457
Report Number1820334-2025-01002
Device Sequence Number12901182
Product Code FFL
UDI-Device Identifier00827002327016
UDI-Public(01)00827002327016(17)280514(10)16601400
Combination Product (Y/N)N
Initial Reporter CountryRO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG32701
Device Catalogue NumberNCT4-024115
Device Lot Number16601400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2025
Supplement Date Received by Manufacturer11/04/2025
Initial Report FDA Received Date08/17/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-