Catalog Number 3549315
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 03/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast reconstruction revision surgery with implantation of a 650cc mentor cpx 4 breast tissue expander with suture tabs.Post-operatively, the patient was diagnosed with a deflated left expander using ultrasound imaging.As a result, the patient underwent exchange with a mentor expander on (b)(6) 2025.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.However, a photo of the suspect medical device was provided.An evaluation using the image was completed.Upon visual evaluation of the image provided in the complaint, the expander was observed deflated.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On december 03, 2025, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Paperwork received with the device provided a patient identifier.Patient identifier: (b)(6).Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On february 15, 2026, mentor completed an evaluation on the returned device.Device evaluation: mentor then conducted a visual inspection, leak testing, and microscopic examination of the device.During the visual analysis of the returned device no apparent damage or visual anomalies were observed.Leak testing was performed, according to the mentor procedure, and it revealed a tear on the anterior view, measuring approximately 0.1 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the tissue expander shell.Based on the information currently available, microscopic examination of the returned product indicates that the tissue expander could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22824524
Report Number1645337-2025-09109
Device Sequence Number17284368
Product Code LCJ
UDI-Device Identifier00081317006817
UDI-Public(01)00081317006817(17)280904(10)9996478
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K130813
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3549315
Device Lot Number9996478
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer12/03/2025
01/28/2026
Initial Report FDA Received Date08/17/2025
Supplement Report FDA Received Date12/24/2025
02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-