• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MORTARA INSTRUMENT ELI380 ERGO WAM WLAN USB DICOM SEC AHA B; ELECTROCARDIOGRAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MORTARA INSTRUMENT ELI380 ERGO WAM WLAN USB DICOM SEC AHA B; ELECTROCARDIOGRAP Back to Search Results
Model Number ELI380-DCS11
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported intermittent wifi connections.There was no adverse event reported.
 
Additional Manufacturer Narrative
The eli 380 is intended to be a high-performance, multichannel resting electrocardiograph.As a resting electrocardiograph, the eli 380 simultaneously acquires data from each lead.Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted.It is a device primarily intended for use in hospitals but may be used in medical clinics and offices of any size.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.Troubleshooting with the customer confirmed the cause of issue was automatic syncing of ecgs.Account turned off feature and issue was resolved.Although there was no injury reported, if the eli380 were to drop wifi connection, it could lead to a serious injury or death.Therefore, baxter is reporting this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELI380 ERGO WAM WLAN USB DICOM SEC AHA B
Common Device Name
ELECTROCARDIOGRAP
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key22825230
Report Number2183461-2025-00010
Device Sequence Number15786910
Product Code DPS
UDI-Device Identifier00812345026917
UDI-Public812345026917
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberELI380-DCS11
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Initial Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-