| Model Number |
SIF-Q180 |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that, during reprocessing, the evis exera ii small intestinal videoscope tested positive for an unknown number of colony forming units (cfus) of sphingomonas paucimobilis.The device was retested on july 9, 2025, and it tested positive for an unknown number of cfus of rhizobium radiobacter.The device was tested again on july 13, 2025, and tested positive for an unknown number of cfus of pseudomonas aeruginosa.The device was tested again on july 22, 2025, and tested positive for an unknown number of cfus of acinobacter iwoffi.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the completed investigation's review of the customer cleaning disinfection and sterilization (cds) processes, results of third-party testing, the device evaluation and the final investigation.The complaint investigation reviewed the customer provided cds processes where information to judge deviation from the ifu was not identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2025, sampling from: all channels, cfu: <1cfu, bacterial identification: bacillus cereus.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|