PHILIPS ULTRASOUND, INC EPIQ CVX; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ CVX DIAGNOSTIC ULTRASOUND SYSTEM
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported the swivel mechanism of their epiq cvx ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.A philips service engineer arrived on site and could not verify the issue.The philips service engineer replaced the bearing and solenoid, and the issue was resolved.A thorough engineering investigation has been performed to identify the root cause of the reported issue.The engineering team determined the failure was a hardware issue in the articulating arm latching mechanism preventing the locking solenoid from fully engaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVX
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
roxanne ramirez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key22827213
Report Number3019216-2025-000285
Device Sequence Number9522963
Product Code IYN
UDI-Device Identifier00884838123960
UDI-Public(01)00884838123960
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ CVX DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795231
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Initial Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-