• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT Back to Search Results
Catalog Number 778426
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a patient underwent ureteroscopy for kidney stones and required removal of a stent.The ureteral stent 778426 had been in place for 30 days.During preparation for removal, it was found that the stent had crusted over, making it impossible to remove smoothly.After clinical assessment, surgical intervention was chosen to remove the stent, and the procedure was completed successfully with the stent removed.It increases surgical procedure time and costs.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a patient underwent ureteroscopy for kidney stones and required removal of a stent.The ureteral stent 778426 had been in place for 30 days.During preparation for removal, it was found that the stent had crusted over, making it impossible to remove smoothly.After clinical assessment, surgical intervention was chosen to remove the stent, and the procedure was completed successfully with the stent removed.It increases surgical procedure time and costs.
 
Additional Manufacturer Narrative
The reported event was confirmed - cause unknown.Received 1 ureteral stent.Verified material number 778426 and batch number ngjs0323.Visual inspection noted a white substance on the stent.The reported event is addressed within the labeling.A review of the device history records did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: b,d,e,h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key22827876
Report Number1018233-2025-06820
Device Sequence Number9336703
Product Code FAD
UDI-Device Identifier00801741014765
UDI-Public(01)00801741014765(11)240521(17)290331(10)NGJS0323
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number778426
Device Lot NumberNGJS0323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Supplement Date Received by Manufacturer02/24/2026
Initial Report FDA Received Date08/18/2025
Supplement Report FDA Received Date03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-