| Catalog Number |
778426 |
| Medical Device Problem Code |
Calcified (1077)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a patient underwent ureteroscopy for kidney stones and required removal of a stent.The ureteral stent 778426 had been in place for 30 days.During preparation for removal, it was found that the stent had crusted over, making it impossible to remove smoothly.After clinical assessment, surgical intervention was chosen to remove the stent, and the procedure was completed successfully with the stent removed.It increases surgical procedure time and costs.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a patient underwent ureteroscopy for kidney stones and required removal of a stent.The ureteral stent 778426 had been in place for 30 days.During preparation for removal, it was found that the stent had crusted over, making it impossible to remove smoothly.After clinical assessment, surgical intervention was chosen to remove the stent, and the procedure was completed successfully with the stent removed.It increases surgical procedure time and costs.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed - cause unknown.Received 1 ureteral stent.Verified material number 778426 and batch number ngjs0323.Visual inspection noted a white substance on the stent.The reported event is addressed within the labeling.A review of the device history records did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: b,d,e,h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|