Catalog Number 8065710001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that during cataract extraction with intraocular lens placement surgery, when placing the sutures into patient eye to secure the sulcus lens the four sutures from same lot kept breaking resulted in anterior vitrectomy.The procedure was completed by using an alternate suture.No patient harm was reported.
 
Additional Manufacturer Narrative
Two unopened sutures, in pouches and blisters, in a bag were received for the report of sutures failed during surgery.Samples were visually inspected and found to be conforming.A dimensional inspection was done for suture diameter, and samples were found to be conforming.Functional testing for suture strength was performed, and samples were found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The returned unopened samples were found to be visually, dimensionally and functionally conforming, therefore the suture failed during surgery as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.However, since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), NYLON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22828146
Report Number2523835-2025-00875
Device Sequence Number13153371
Product Code GAR
UDI-Device Identifier00380657100019
UDI-Public(01)00380657100019(17)271031(11)221103(10)155LL3
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K930517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065710001
Device Lot Number155LL3
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer11/06/2025
Initial Report FDA Received Date08/18/2025
Supplement Report FDA Received Date12/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-