5 of 6 reports.Other mfg report numbers: 3013886523-2025-00191.2648988-2025-00028.3013886523-2025-00192.2648988-2025-00029 2648988-2025-00035.This case was opened as the customer estimated around 50 times bag dysfunction occurred.This issue occurred in the icu & neuro departments of the facility.Devices: 821730c (eds3 drainage system); ins8700 (integra external csf drainage replacement bags).Date of event: beginning june 9th 2025 (during several weeks) a facility reported: - increased flow rate when the valve is opened to empty the chamber.- air rising in the catheter - hyper-drainage - slow emptying - air appearing in the patient's ventricle (visible on ct scan) immediate action(s) taken: change of equipment, discontinue use.Additional information subsequently received as follows: - how many times has this malfunction been observed? answer: about fifty bag malfunctions and air rising due to slow emptying.- on what dates? answer: for a year or more, because at the beginning they noticed very slow emptying (several minutes to empty a bag so some nurses went to see other patients while waiting (but be careful: risk of forgetting) or others use a syringe to speed up emptying in accordance with the rules of asepsis (but risk of aggravating hyper-drainage, and forgetting a syringe), but did not think it was a problem with the md but rather a handling problem, but after several internal observations, they finally reported it to their head of department at the end of (b)(6) 2025.- for each malfunction, what are the batch numbers of the 821730c and ins8700 devices used? answer: i cannot answer this question, because the batches used were unfortunately not kept.- for each malfunction, what symptoms did the patient experience that indicated an over-drainage problem? answer: headaches, air in the ventricles (like the ct scan photos i sent you at the beginning of july).- for each malfunction, how is the patient doing following the system change? answer: there is no objective feedback because the care team often decides to either remove the evd, switch to another shunt system, or transfer the patient to another hospital for continuity of care.- for each malfunction, is the patient elderly, an adult, or a child? answer: adult.- how many devices are available to be returned for investigation? answer: 0 unfortunately - could you please clarify the box that states "increased flow rate" when opening the valve to empty the chamber: answer: the dropper speeds up (more drops dripping).- air rising in the catheter: answer: air rises in the tubing and pushes the fluid not towards the chamber but back into the brain.- hyper-drainage: answer: excessive elimination of csf - slow emptying: answer: drainage sometimes blocked, or flowing over several minutes.- appearance of air in the patient's ventricle (visible on ct scan)"? does the burette empty slowly into the drainage bag; and (at the same time) the csf flow from the patient's line to the burette is increased? answer: yes, but it is difficult to quantify.- what is the procedure when using syringes to accelerate emptying? answer: a procedure doesn't exist; it's performed according to aseptic rules, but it's actually a hack.- was the malfunction observed after the patient was transported? could the device have been placed horizontally at any time, thus wetting the hydrophobic membrane located above the drip area? answer: no, observed even before, according to the nurses.They know that before placing it horizontally, the chamber must be emptied so that the hydrophobic membrane is intact.Unfortunately, even when intact, there was a negative pressure problem that drained at the same time as the emptying.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The eds3 drainage system was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined, however, the possible root cause for the reported complaint could be due the incompatibility between the collection bag used (belongs to the accudrain evd system) and the eds3 system.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|