|
ETHICON INC. SFXSPI PDS+ UNI VIO 18IN 3-0 SA PS-2 PR; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
|
Back to Search Results |
|
| Catalog Number |
SXPP1B105 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 8/18/2025.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: when did the needle detach or pull off from the suture: detached inside the package, during removal from the package, during handling prior to use on the patient, or during use on the patient? unk.Please provide the title of external person providing answers to follow-up (e.G.Nurse, surgeon, risk manager) (b)(6).Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at sukagawa and will be shipped.Please check rmao.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During an unknown surgery, the suture was detached from the needle easily.It was not a control release needle.Another product was used to complete the case.There were no adverse consequences to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/16/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d9, h3, h6 a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with six unopened samples were received for analysis.Product code sxpp1b105.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|