Model Number 3300TFX27MM
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 10/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11.Additional narrative: surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as its availability is unknown.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was learned via implant patient registry that a 3300tfx 27mm aortic valve was explanted and replaced after an implant duration of 9 years due to calcification.The patient presented with sob.The replacement valve was unknown.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11.Additional narrative updated h6.Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years and hyperlipidemia.A device history record (dhr) review is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
one edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22830619
Report Number2015691-2025-06767
Device Sequence Number4464044
Product Code DYE
UDI-Device Identifier00690103176162
UDI-Public(01)00690103176162(17)150729
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/29/2015
Device Model Number3300TFX27MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Supplement Date Received by Manufacturer09/11/2025
Initial Report FDA Received Date08/18/2025
Supplement Report FDA Received Date09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age39 YR
Patient SexMale
-
-