| Model Number |
KD-V411M-0725 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the single use 3-lumen sphincterotome v's wire fractured.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography and was completed with an olympus back-up device.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: h2, h3, h6, h11.The device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.An electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Olympus will continue to monitor the performance of this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction is being made to previously submitted h4 - device manufactured date, which was transposed on the previous report.The correct date is 8/8/24.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|