Model Number KD-V411M-0725
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported the single use 3-lumen sphincterotome v's wire fractured.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography and was completed with an olympus back-up device.There were no reports of patient harm.
 
Event or Problem Description
No additional information received from the customer.
 
Additional Manufacturer Narrative
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: h2, h3, h6, h11.The device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.An electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Olympus will continue to monitor the performance of this device.
 
Additional Manufacturer Narrative
Correction is being made to previously submitted h4 - device manufactured date, which was transposed on the previous report.The correct date is 8/8/24.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22830745
Report Number9614641-2025-01387
Device Sequence Number19977873
Product Code KNS
UDI-Device Identifier04953170184031
UDI-Public04953170184031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/22/2025
Supplement Date Received by Manufacturer08/27/2025
09/16/2025
Initial Report FDA Received Date08/18/2025
Supplement Report FDA Received Date09/12/2025
09/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-