• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P; No Match Back to Search Results
Model Number PM3262
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.The device was received with normal telemetry and output.Visual inspection of the header attachment area detected a bonding anomaly between the epoxy header and titanium case.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The device longevity assessment found the device was operating above the elective replacement indicator (eri) voltage, indicating normal usage.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.
 
Event or Problem Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP RF CRT-P
Common Device Name
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22831085
Report Number2017865-2025-97947
Device Sequence Number14593737
Product Code NIK
UDI-Device Identifier05414734509091
UDI-Public(01)05414734509091(10)A000075254(17)200831
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2020
Device Model NumberPM3262
Device Lot NumberA000075254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Initial Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-