| Catalog Number |
SRF-240 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
04/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "capsular contracture baker grade iv".
|
| |
|
Event or Problem Description
|
|
Patient reported " capsular contracture with unknown baker grade" via "mammogram".Later healthcare professional reported "capsular contracture baker grade iv".This record is for the "left" side.The device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Lab analysis: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe as it is not related to the manufacturing process.As per the investigation procedure, deformation and creases were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Patient reported " capsular contracture with unknown baker grade" via "mammogram".Later healthcare professional reported "capsular contracture baker grade iv".This record is for the "left" side.The device was explanted and replaced.
|
| |
|
Search Alerts/Recalls
|