• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER MAXGRIP REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER MAXGRIP REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 705.13P
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned instrument was received in its inner blister covered with the tyvek lid foil showing the lot information.The forceps was returned in a closed state, and it shows signs of surgery residues.The returned instrument was visually inspected with the aid of a photomicroscope using various magnifications.It was noticed that the tube is slightly bent.The device was then functionally tested, and no actuation of the forceps is possible.The actuation mechanism itself is still in working condition, but the forceps stuck in a close position.The forceps can only get opened again by manually pushing down the metal shaft, indicating that the bonding connection between the shaft and stiffening sleeve got broken.As the blister was opened by the customer, the product was handled.Therefore, the point in time when the malfunction occurred can no longer be determined, nor can the strain put on the device be reconstructed.The customer¿s complaint is confirmed, but a root cause for the connection failure cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.It cannot be determined how or where the damage to the sample occurred.Therefore, the root cause for the customer's reported event could not be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that ophthalmic forceps tip did not open.The timing of the surgery was unknown.The surgery was completed after replacing the product with another one.The involved eye and the patient identifier are not available.No patient impact was reported.Additional information has been requested; none received till date.
 
Additional Manufacturer Narrative
Additional information provided in sections b.5.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information received that the event occurred during surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER MAXGRIP REVOLUTION DSP FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22839194
Report Number3003398873-2025-00354
Device Sequence Number21187939
Product Code HNR
UDI-Device Identifier07612717070721
UDI-Public(01)07612717070721(17)271031(11)241122(10)175L4Y
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705.13P
Device Lot Number175L4Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/27/2025
Supplement Date Received by Manufacturer08/20/2025
Initial Report FDA Received Date08/19/2025
Supplement Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-