| Catalog Number |
221261 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported before and during use of bd bbl¿ trypticase¿ soy agar with 5% sheep blood (tsa ii), white fungal contamination was observed.Uninoculated plates, plates used for non-patient testing, and an unknown number of plates used with an unspecified number of patient samples were involved.No confirmatory testing was performed and the fungal contaminate was not identified.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported before and during use of bd bbl¿ trypticase¿ soy agar with 5% sheep blood (tsa ii), white fungal contamination was observed.Uninoculated plates, plates used for non patient testing, and an unknown number of plates used with an unspecified number of patient samples were involved.No confirmatory testing was performed and the fungal contaminate was not identified.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: during manufacturing of material 221261, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record review for batch 5114200 was satisfactory and no quality notifications were generated during manufacturing and inspection.A review of quality notifications and complaint data for the manufacturing plant, which includes material 221261, revealed no batches or materials exhibiting contamination trends during calendar year 2025.Batch 5114200 was manufactured in april of 2025, and no contamination-related complaint trends have been reported.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 5114200 for this defect.There were no retention samples available for investigation of this complaint.A total of five photos were submitted by the customer to support their complaint.The observations from each photo are as follows: three photos show a sleeve from 221261/5114200, sleeve label is visible.The top plate appears to be contaminated.One photo shows a sleeve from 221261/5114200, sleeve label is visible.One photo show two plates one upside and one downside, contamination can be seen on the surface of the plate.No returns were provided for this investigation.This complaint can be confirmed based on the photos provided.As part of the investigation manufacturing record including equipment logs, cleaning documentation, and environmental monitorin related to the production of batch 5114200 was reviewed.No definitive root cause was identified for the contamination observed in this batch.While the root cause remains undetermined, the contamination event was confirmed to be isolated to products shipped to baylor scott & white research institute.Based on the current findings no corrective actions are identified at this time.Bd will continue to monitor and trend contamination-related complaints to ensure ongoing product quality.
|
| |
|