| Catalog Number |
368861 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® k2e (edta) 7.2mg plus blood collection tubes, foreign matter was found in one (1) tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd received four photos for investigation.Evaluation of these photos indicated a yellowish edta clump in the tube, recognized as an aesthetic defect with no impact on the efficiency of the tube.Additionally, 100 retained samples were visually inspected, and edta clumps were found.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter - other.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® k2e (edta) 7.2mg plus blood collection tubes, foreign matter was found in one (1) tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|