Model Number CC60WF
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, while trying to load the iol into the cartridge, it was noticed that the lens appeared to have a defect.The iol was handed to the surgeon, not in a cartridge, just the iol to view under the microscope and the surgeon confirmed that the iol was scratched.There was no patient contact.The surgery was completed on same day with backup iol.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22846410
Report Number1119421-2025-02286
Device Sequence Number8403611
Product Code HQL
UDI-Device Identifier00380652396790
UDI-Public(01)00380652396790(17)290223(11)250225(10)15999119
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCC60WF
Device Lot Number15999119
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2025
Initial Report FDA Received Date08/19/2025
Date Device Manufactured02/25/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II IOL DELIVERY SYSTEM, CARTRIDGE A; MONARCH II LOADING FORCEPS; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-