ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR KNEE, REVISION FEMUR, 7 RIGHT Back to Search Results
Catalog Number 245-02-107
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 07/23/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to periprosthetic fracture.
 
Additional Manufacturer Narrative
The agent reported "(periprosthetic fracture revision to distal femur replacement)".The previous surgery and the surgery detailed in this event occurred 2 days apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.First, there was a nonconformance associated with the main part # 348-12-707 empowrrevisionkneetm revvvc+ tibialinsert e+, sz7, 12mm, which documents that out of 40 parts lot, 2 items were rejected and scrapped during final pack quality check.Second, there was a nonconformance associated with the main part # 354-15-050 empowr knee, stem extender,15mm x 50mm, which documents that out of 40 parts lot, 1 item was rejected and scrapped during metal finishing - #8 glass bed blast per engineering drawing.All other items in the lot were met with fit, form and function.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to periprosthetic fracture.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR KNEE, REVISION FEMUR, 7 RIGHT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key22846529
Report Number1644408-2025-01150
Device Sequence Number4419351
Product Code JWH
UDI-Device Identifier00190446681090
UDI-Public00190446681090
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K213793
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number245-02-107
Device Lot Number2690A1012
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2025
Initial Report FDA Received Date08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
130-03-729 LOT: 154N1899; 348-12-707 LOT: 4638A1001.; 352-04-107 LOT: 329V1105.; 354-14-040 LOT: 107Y1167.; 354-14-065 LOT: 108Y1119.; 354-15-025 LOT: 2735A1036.; 354-15-050 LOT: 2736A1028.
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age84 YR
Patient SexFemale
-
-