| Model Number |
8404EXC |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Optical Problem (3001)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is not yet completed, investigation results are pending.This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported the light is very low and image is dark.No patient or procedure involvement.No negative impact in state of health reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's evaluation results: the device was sent to the manufacturer for inspection.During the manufacturer's technical inspection, the error description "the light is very low and image is dark" was confirmed, and further defects were detected.The device has not been repaired or maintained until now.The device passed the quality check at the time of delivery.Based on the defects found during the technical inspection it is concluded that the most likely cause for the described error is the wear of the adhesive at the tip of the c-mac blade, coming from wear and tear during use and the reprocessing process.The wear of the adhesive causes liquid to penetrate the blade, which affects the electronics and causes the light is weak.This event is filed under internal complaint id (b)(4).
|
| |
|