Model Number 20014
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a coronary artery bypass graft procedure, it was reported that two dlp aortic root cannulae with vent lines failed.The cannulae remained non-functional after activation.The first cannula, sourced from a third-party surgical pack, was inserted into the aorta, the stylet was removed, and the pump suction line was attached to the vent line, but the vent did not function.Visual inspection confirmed no kinks or obstructions in the tubing, and independent testing verified the pump suction line was working.After reattaching the line, the vent remained non-functional, leading to removal and replacement with a single-packed dlp aortic root cannula, which also failed in the same manner.Both cannulae were removed, and an alternative aortic root cannula was used successfully.No patient complications have been reported as a result of this event.Medtronic received additional information that a flow issue was observed.It appeared that there was no blood flow through the cannulae.The bypass time was prolonged by 3 to 5 minutes as a result of the event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Device evaluation summary: visual inspection shows no outward signs of any damage.A syringe filled with de-ionized water was connected to the device and when it was engaged there was flow observed through the tip and the vent line.Reason for the return was not confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DLP AORTIC ROOT ANTEGRADE ANTEGRADE CANNULA WITH VENT LINE
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key22849183
Report Number2184009-2025-00977
Device Sequence Number8457768
Product Code DWF
UDI-Device Identifier00613994495549
UDI-Public(01)00613994495549(17)280203(10)0230553739
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K810548
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20014
Device Catalogue Number20014
Device Lot Number0230553739
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2025
Supplement Date Received by Manufacturer09/19/2025
Initial Report FDA Received Date08/20/2025
Supplement Report FDA Received Date10/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-