Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.One open box with thirty representative unopened samples received for analysis.Product code pdp311h.Following the sampling plan, a visual inspection was conducted on thirteen samples.The packets were open, and the swage and attachment area were noted to be as expected.These ones were tested to verify the dispensability of the sutures, and it was not possible in all samples due to the suture being found hard to release from the tray by applying excessive force.Due to this inconsistency, the remaining twenty-three samples were tested to verify the dispensability of the sutures, and it was not possible in all samples due to the suture being found hard to release from the tray by applying excessive force.Based on the information currently available, the indispensable suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|