MEDOS INTERNATIONAL SàRL RGDLOOP ADJUSTABLE XLARG; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 232449
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported that during an anterior cruciate ligament (acl) repair surgical procedure it was observed that the rgdloop adjustable xlarg device was broken.Another like device was used to complete the procedure successfully.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.¿.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device, and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the device comes in used condition.The white suture is broken and frayed.The green suture was not returned.No other anomalies were noted.An investigation was made with the manufacturer; as a result, the white suture is utilized to assembly onto a graft construction which provides fixation.Since the suture breakage couldn't be originated during manufacturing process, the breakage can be attributed to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.As per ifu the steps for a proper insertion are provided.The overall complaint was confirmed as the observed condition of the rgdloop adjustable xlarg would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.It has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the device comes in used condition.The white suture is broken and frayed.The green suture was not returned.No other anomalies were noted.An investigation was made with the manufacturer; as a result, the white suture is utilized to assembly onto a graft construction which provides fixation.Since the suture breakage couldn't be originated during manufacturing process, the breakage can be attributed to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.As per ifu the steps for a proper insertion are provided.The overall complaint was confirmed as the observed condition of the rgdloop adjustable stnd would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.It has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RGDLOOP ADJUSTABLE XLARG
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SàRL
chemin blanc 38
le locle CH-02 400
SZ   CH-02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22850522
Report Number1221934-2025-03201
Device Sequence Number8373348
Product Code MBI
UDI-Device Identifier10886705024117
UDI-Public(01)10886705024117(10)128L959(17)251231
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140324
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue Number232449
Device Lot Number128L959
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2025
Supplement Date Received by Manufacturer09/01/2025
09/23/2025
09/29/2025
Initial Report FDA Received Date08/20/2025
Supplement Report FDA Received Date09/02/2025
09/26/2025
10/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-