• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL COBALT¿ HF CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL COBALT¿ HF CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTPB2D1
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Codes Tachycardia (2095); Loss of consciousness (2418); Shock from Patient Lead(s) (3162)
Date of Event 07/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The patient was seen by their following physician and it was determined that the shock the patient received was appropriate.
 
Event or Problem Description
It was reported by the patient that they received a shock from the cardiac resynchronization therapy defibrillator (crt-d) and passed out.The patient experienced multiple instances of ventricular tachycardia (vt) and noted they were really sick.The crt-d remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT¿ HF CRT-D MRI SURESCAN¿
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22851558
Report Number9614453-2025-03288
Device Sequence Number4488802
Product Code NIK
UDI-Device Identifier00763000178222
UDI-Public(01)00763000178222(17)250528(20)02
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/28/2025
Device Model NumberDTPB2D1
Device Catalogue NumberDTPB2D1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Supplement Date Received by Manufacturer08/29/2025
Initial Report FDA Received Date08/20/2025
Supplement Report FDA Received Date09/26/2025
Date Device Manufactured12/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4968-60 LEAD, 6937A-58 LEAD, 4968-35 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age73 YR
Patient SexMale
-
-