| Model Number |
LF1944 |
| Medical Device Problem Code |
Energy Output Problem (1431)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure the device was not sealing, there was no evidence of deliver energy and end tones were heard.Another ligasure device was used successfully with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a bariatric bypass, the device was not sealing, there was no regrasp alert, no evidence of deliver energy and end tones were heard.The device was cleaned when eschar was evident on the jaws.Another device was used successfully with the same generator.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the seal plate was damage, consistent with arcing.Functional testing found that the damage to the seal plate(s) is consistent with the user inadvertently closing jaws on a hard/metallic object and activating the device.Damage to the seal plate can result in alarm activations and difficulty with activating the device.The product analysis found the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The blade moved smoothly along the knife track and returned to the home position when the trigger was released.The knife cut of the device was tested on a silicone test strip with acceptable results.The heel gap of the device was measured and it was found to be within specification.The device was activated multiple times on a saline soaked gauze pad but had re-grasp alarms.The seal cycle was interrupted.No electrical or wiring issues were found.Manufacturing does 100% inspection during the assembly and packaging process.The defective sample would have been rejected if found prior to shipping, if this was an assembly defect.It was reported that the de vice was not sealing.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: of a damaged seal plate was observed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not attempt to seal or cut over clips or staples as incomplete seals will be formed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a bariatric bypass, the device was not sealing, there was no regrasp alert, no evidence of deliver energy and end tones were heard.The device was cleaned when eschar was evident on the jaws.Another ligasure device was used successfully with the same generator.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|