On (b)(6) 2025, a sales representative reported via phone that an ar-3636h suture anchor, and (2) ar-16992n scorpion¿ needles failed.This occurred during a hip labral repair on (b)(6) 2025 when the ar-3636h repair stitch did not pass through the anchor.The device was removed and another ar-3636h was used.The (2) ar-16992n broke during the first pass.All fragments were retrieved.The case continued using an ar-4068hl.There was about a 10-minute delay where additional anesthesia was administered.The representative was unsure but said he could guess the bone quality was hard.
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during firing.Dfu-0392-sub-eo: avoid striking bone or using excessive force to pass the needle through fibrous/calcific tissue.If excessive force is encountered, replace the needle, reposition the device and pass in a different area.The complaint allegation could not be confirmed as the device was not received for investigation.
|