• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE Back to Search Results
Model Number 7300TFX
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 07/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 25mm 7300tfx mitral valve underwent a valve-in-valve procedure after an implant duration of 8 years, 4 months due to severe mitral stenosis and mitral regurgitation.The patient presented with heart failure.The tmvr was successfully performed with a 26mm 9755rsl valve.The patient was stable post-procedure and was discharged on pod #1.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: updated sections: d4 (expiration date), g3, g6, h2, h4, h6 (type of investigation, investigation findings, and investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including coronary artery disease and liver disease.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22852311
Report Number2015691-2025-06843
Device Sequence Number14557654
Product Code DYE
UDI-Device Identifier00690103186253
UDI-Public(01)00690103186253(17)200210
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/10/2020
Device Model Number7300TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer09/15/2025
Initial Report FDA Received Date08/20/2025
Supplement Report FDA Received Date09/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/11/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age50 YR
Patient SexFemale
-
-