Model Number DSX900S11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of foam degradation.There is no allegation of serious or permanent harm or injury.No medical intervention was specified as required by the patient.During the evaluation of the device, at the manufacture's service center the device was visually inspected and found evidence of foam degradation inside assembly.The device has been scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key22852342
Report Number2518422-2025-109252
Device Sequence Number4362155
Product Code MNS
UDI-Device Identifier00606959424388
UDI-Public(01)00606959424388
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K090539
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 08/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSX900S11
Device Catalogue NumberDSX900S11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2025
Initial Report FDA Received Date08/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-