A health care professional reported that during an intraocular lens (iol) implant procedure, there was an implantation failure, the leading haptic does not fold.The incident occurred during the advancement of the implant in the injection system.The device came into contact with the patient eye.The lens was replaced with a similar one.Additional information has been requested.
|
|
The lens was returned positioned incorrectly in the base component of the lens case, placed inside an opened, folded pouch in the lens carton.The lens case lid with mylar was not returned.Solution was dried on the lens.One haptic was bent in the gusset area.The lens was cleaned with klrp (klotho-related protein) to further evaluate.A small scuff mark was observed on the anterior optic surface near the edge.The bent haptic relaxed into the original position after removal of the dried solution.Both haptics were easily pliable using tweezers.A fold test was conducted using folding tweezers.The lens folded and unfolded with no abnormalities observed.A device history record review and a non-conformance-based review of the lot was performed and did not reveal any potential contributing factors to the reported complaint.All corresponding production release specifications defined in the device master record were met.Associated product information was not provided.It is unknown if qualified associated products were used.The product investigation could not determine a root cause for the complaint.The sample was returned with evidence of customer handling, an optic scuff mark, and a bent haptic.The lens was cleaned to remove the dried solution.The bent haptic relaxed to the original position.Both haptics were easily pliable using tweezers.A fold test was conducted using folding tweezers.The lens folded and unfolded with no abnormalities observed.It is unknown if qualified associated products were used.The instructions for use (ifu) instructs: company foldable intra ocular lens (iols) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and or damage.It is unknown if adequate viscoelastic was used in the cartridge.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturers recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|