• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number DDU-A2450
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that their aed's asi is flashing red and reporting a service code.They reported that this did not occur during patient use.
 
Additional Manufacturer Narrative
Although the customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting.Further analysis indicated that the battery life expectancy was reduced as a result of the customer's failure to promptly address repeated red asi warnings.As a follow up with the customer, the proper maintenance of this device was reviewed.Although the original customer report did not involve patient use and was not considered reportable, evaluation confirmed a malfunction that could delay or prevent therapy delivery in a clinical situation.As such, this event is being reported in accordance with 21 cfr 803.50.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key22854087
Report Number3003521780-2025-01136
Device Sequence Number4419285
Product Code MKJ
UDI-Device Identifier10815098020284
UDI-Public(01)10815098020284
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-A2450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Initial Report FDA Received Date08/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1