|
Issue identified during product analysis.H3 device evaluation summary: medtronic conducted an investigation based on all information received.It was reported that the pb980 ventilator had a system error.The device was available for evaluation.The service personnel (sp) inspected the ventilator, found the unit displayed multiple error messages and replaced the backup ventilation (buv) solenoid, safety valve, direct current - direct current (dc-dc) converter printed circuit board assembly (pcba), and oxygen (o2) sensor.Additionally, the unit failed the battery test portion of extended self test(est), so sp replaced the breath delivery(bd) power controller/distribution pcbas.Sp performed 10k preventive maintenance.The unit passed all calibrations and tests as per the manufacturer's specifications at the time of service.The bd power controller/distribution pcbas were not returned to medtronic for analysis.One oxygen (o2) sensor was returned to medtronic for analysis.On visual inspection, it was observed that the o2 sensor was a non-mdt part.The non-mdt o2 sensor was not installed into the ventilator by a medtronic field service engineer, but could have been installed by the customer.The buv solenoid was returned to medtronic for analysis.The component was visually inspected and functionally tested with no malfunction or product deficiency observed.One safety valve was returned to medtronic for analysis.Visual inspection of the safety valve found damage to the connector.Although the connector was damaged, the safety valve was functionally tested with no malfunction or product deficiency observed.The cause of the damage could not be determined.One dc-dc converter pcba was returned to medtronic for analysis.Visual inspection of the dc-dc converter pcba found no anomalies.The dc-dc converter pcba was attached to the test ventilator and powered up.While performing calibration, it was observed that the test ventilator was making a very loud noise and the safety valve was turning on and off.The fault was isolated to the switching regulator (u2) on the dc-dc converter pcba.If a failure of the u2 occurs while in use on a patient, the ventilator response would be a loss of ventilation to the patient.The cause of the reported event was isolated to faulty component u2 on the dc-dc converter pcba.The cause of the battery test portion failure during est was isolated in the field to a potential fault in the bd power controller/distribution pcbas.The events are included in a trending and monitoring plan.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|