| Model Number |
E161 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.Additional information was received that this device was explanted due to normal battery depletion (nbd).
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.Additional information was received that this device was explanted due to normal battery depletion (nbd).
|
| |
|