BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number E161
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.Additional information was received that this device was explanted due to normal battery depletion (nbd).
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a product performance issue or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a performance issue or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Additional Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range shock impedance measurements on the right ventricular (rv) lead.Technical services (ts) confirmed this has not been a gradual rise and that there is a high degree of variability up and down.Reprogramming was performed.No adverse patient effects were reported.This device remains in service.Additional information was received that this device was explanted due to normal battery depletion (nbd).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22856375
Report Number2124215-2025-58414
Device Sequence Number18462995
Product Code LWP
UDI-Device Identifier00802526480768
UDI-Public(01)00802526480768(17)141010
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/10/2014
Device Model NumberE161
Device Catalogue NumberE161
Device Lot Number103160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2025
Supplement Date Received by Manufacturer10/15/2025
12/31/2025
Initial Report FDA Received Date08/20/2025
Supplement Report FDA Received Date10/23/2025
01/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/16/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexMale
-
-