| Catalog Number |
206967 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Dark gray particles were found inside the hip end effector after surgical procedure using mako robot.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding foreign matter involving a mako end effector was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection did not confirm the event.The device did not have dark gray particles inside.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there are no other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Event or Problem Description
|
|
No new information available.
|
| |
|