| Model Number |
406840 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One 8f fast-cath introducer sheath and dilator were received for evaluation.Resistance was noted near the distal end of the dilator when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and a build-up of skived material was noted at the distal end of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the needle insertion difficulty is consistent with skiving.The cause of the skiving remains unknown.
|
| |
|
Event or Problem Description
|
|
This report is to advise of skiving that was noted during analysis.
|
| |
|
Search Alerts/Recalls
|