E1: customer (person): phone: (b)(6).H3: device evaluated by mfg? device evaluation is anticipated but has not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
It was reported that upon opening the package of a n-compass nitinol stone extractor, the "holder of the sheath" was discovered to be broken.The device was not used, and another device was obtained to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Additional information: d9 - date returned to mfg.Correction: d4 - primary udi number.Investigation ¿ evaluation: it was reported that upon opening the package of a n-compass nitinol stone extractor, the "holder of the sheath" was discovered to be broken.The device was not used, and another device was obtained to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures, as well as a visual inspection and functional test of the returned device, were conducted during the investigation.One n-compass nitinol stone extractor was returned in an open package with label.Upon inspection, there was significant damage to the basket sheath.The basket sheath was completely broken approximately midway down the device.The support tubing was cracked at the handle.Basket formation could not be opened.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot did not reveal any related quality control discrepancies.A complaint history search identified no other events reported for the related failure mode.Based on the available information, cook has concluded that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The product ifu, t _ ntse_ rev1, provides the following information to the user: precaution: do not use excessive force to manipulate this device.Damage to the device may occur.The evidence from the complaint file, dhr, complaint history, and quality control documents indicates that the complaint device was manufactured to specification as well as other items in the lot or similar devices in the field or in house.Based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded a cause for the damage to the sheath cannot be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|