MAUDE Adverse Event Report: ZIMMER BIOMET, INC. AFFIXUS HIP FRACTURE NAIL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. AFFIXUS HIP FRACTURE NAIL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 814611380
Medical Device Problem Codes Fracture (1260); Positioning Problem (3009)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/28/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a procedure, the ar drill misfired, hitting the nail and causing the drill bit to break.When drilling for the ar screw, the blue ar sheath was flush against the jig but the drill bit was not going through the hole in the nail.As a result, it was hitting the nail and misfiring.The sleeve was then removed, triple blue sleeve replaced and attempted to drill again but drill bit still hitting the nail and then broke off inside the patient.The tip of the ar drill was buried inside the greater troch and was unable to be removed.As a result, hardware retained.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: item# 211201201; lot# cx2785m01.H6: proposed component code - mechanical (g04)- drill.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d1; d2; d4; d6a; d9; g1; g3; g4; g6; h1; h2; h4; h6.H6: proposed component code, mechanical (g04) nail.Customer has indicated that the nail will not be returned to zimmer biomet for investigation, as the product was implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d4; d9; g1; g3; g6; h1; h2; h4.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d4; d9; g1; g3; g6; h1; h2; h3; h6; h10.No product was returned or pictures provided or the nail; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the tip of a drill bit fragment is partially visualized in the proximal femoral metaphysis.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event for part 814611380.A definitive root cause cannot be determined.The event was confirmed.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFIXUS HIP FRACTURE NAIL
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22860884
Report Number0001825034-2025-02606
Device Sequence Number7779363
Product Code HSB
UDI-Device Identifier00887868567254
UDI-Public(01)00887868567254(17)350410(10)67088223
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K100238
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number814611380
Device Lot Number67088223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2025
Supplement Date Received by Manufacturer01/26/2026
02/06/2026
03/24/2026
Initial Report FDA Received Date08/21/2025
Supplement Report FDA Received Date02/06/2026
02/11/2026
03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H11.
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-