| Catalog Number |
ARS743000 |
| Medical Device Problem Codes |
Break (1069); Degraded (1153); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The instrument broke during surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: please refer to h6 device code.The reported event could be confirmed since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following, the device received shows signs of extensive usage evidenced by the appearance of scratches wear marks water marks and severe discoloration.Device also has corrosion marks near the identification engravings or etching.Broken piece of the rod is stuck inside the 20 degree hole of the inserter extractor handle.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation the root cause was attributed to a combination of wear and user related issue.The failure was due to frequent and improper usage.If any further information is provided the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
The instrument broke during surgery.
|
| |
|