It was reported that one week post operation, the ct lucia 602 iol (intraocular lens) had rotated and affected the patient's vision.The patient elected to return and do an iol exchange for another lens.The iol was explanted on (b)(6) 2025.The surgical technique used was not reported.The z28 cartridge was utilized (injector unspecified), with no noted lens damage or patient complications.
|
|
The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lens that would have caused or contributed to the nature of this complaint.Additionally, our lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operating procedures and inspections and have met all criteria for release.The device is not available for investigation.Thus, a proper device analysis could not be completed nor the reported issue be confirmed.At this time, without a proper device analysis, a definitive root cause cannot be determined.However, there is no indication of a product quality issue with respect to manufacturing, design, or labeling of the lens.Based on the information provided, the risk assessment for the lucia product, and our experience we have determined that the following factors may have caused or contributed to the reported issue is but is not limited to: lens placement technique, loading strategy, accessory device support poor handling during folding and inserting.
|