MAUDE Adverse Event Report: ABBOTT MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; Introducer, catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; Introducer, catheter Back to Search Results
Model Number 406849
Medical Device Problem Codes Flaked (1246); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
This report is to advise of skiving found in the sheath during analysis.
 
Additional Manufacturer Narrative
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.Resistance was noted when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and a build-up of skived material was noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ artmt600108998 ver.B fast-cath transseptal guiding introducer instructions for use (ifu) states, ¿during insertion, always use the stylet to facilitate needle passage through the dilator/sheath assembly.Failure to use the stylet may inhibit advancement of the needle and may result in inadvertent needle puncture of the dilator/sheath assembly or skiving of material from the inner surface of the dilator.¿ the cause of the needle insertion difficulty and skiving is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.
Common Device Name
Introducer, catheter
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key22868879
Report Number3005334138-2025-00514
Device Sequence Number8894619
Product Code DYB
UDI-Device Identifier05414734203388
UDI-Public(01)05414734203388(10)10545178(17)270930
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406849
Device Catalogue Number406849
Device Lot Number10545178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2025
Initial Report FDA Received Date08/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1