| Model Number |
406849 |
| Medical Device Problem Codes |
Flaked (1246); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of skiving found in the sheath during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.Resistance was noted when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and a build-up of skived material was noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ artmt600108998 ver.B fast-cath transseptal guiding introducer instructions for use (ifu) states, ¿during insertion, always use the stylet to facilitate needle passage through the dilator/sheath assembly.Failure to use the stylet may inhibit advancement of the needle and may result in inadvertent needle puncture of the dilator/sheath assembly or skiving of material from the inner surface of the dilator.¿ the cause of the needle insertion difficulty and skiving is consistent with not following the instructions for use.
|
| |
|