COVIDIEN LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF5644
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a procedure; the device¿s locked and would not open.The jaw did not get stuck on tissue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the device's jaw was stuck on eschar buildup.Functional testing found that the build up was cleaned and the device was working properly.More frequent cleaning of the jaws could have prevented build up of eschar.The product analysis found the mechanical function of the device's jaw opening and closing was functioning properly.It was reported that the device¿s locked and did not open.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22870320
Report Number1717344-2025-01082
Device Sequence Number15723705
Product Code GEI
UDI-Device Identifier10884521521742
UDI-Public(01)10884521521742(17)280529(10)31510175X(20)01
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K161804
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF5644
Device Catalogue NumberLF5644
Device Lot Number31510175X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2025
Supplement Date Received by Manufacturer10/13/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-