Model Number 55750026540
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having transcutaneous vertebral body formation/transcutaneous posterior fusion for osteoporotic vertebral fracture.It was reported that during cement fi lling of the cement screw; cement leaked to the screw head.Because cement leaked to the screw head, the cement delivery guide tip was retained.The planned percutaneous surgery ended up not being entirely percutaneous, resulting in a significantly larger incision being made.There was a delay of greater than 60 minutes in the overall procedure time.A total of 8 screws were used in the fixed range th11/l3.Only the right side of l2 leaked cement on the screw head side.The guide tip could be removed, but after inserting the rod, the set screw could not be inserted on the right side of l2.For the right side of l2, the set screw installation was abandoned.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
B2: other - the planned percutaneous surgery ended up not being entirely percutaneous, resulting in a significantly larger incision being made.There was a delay of greater than 60 minutes in the overall procedure time.H6: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® FENESTRATED SCREW SET
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22870449
Report Number2647346-2025-00070
Device Sequence Number9081614
Product Code NKB
UDI-Device Identifier00763000148829
UDI-Public(01)00763000148829(17)300428(10)H6049344
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201362
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55750026540
Device Catalogue Number55750026540
Device Lot NumberH6049344
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2025
Initial Report FDA Received Date08/22/2025
Date Device Manufactured04/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-